OCR A Level Physics A: Gamma camera, PET scanning and ultrasound (H556) – Revision Notes

Condensed H556 revision notes for 6.5.2 and 6.5.3: tracers, gamma camera, PET, ultrasound, acoustic impedance, gel and Doppler, with a self-test.

Subject
Physics
Level
A LEVELS
Topic
Gamma camera, PET scanning and ultrasound
Updated

Aligned to OCR A Level Physics (H556), Version 3.0 (March 2026), for first assessment in 2017. Official specification .

Syllabus page (what it covers and how it is assessed): OCR A Level Physics.

Syllabus points this page covers

H556

  • 6.5 Medical imaging

Found an error? Report a correction.

Need help with this topic? Request a free trial class for A Level Physics (H556).

These revision notes condense sections 6.5.2 (Diagnostic methods in medicine) and 6.5.3 (Using ultrasound) of the OCR Advanced GCE in Physics A (H556) specification, Version 3.0 (March 2026), for first assessment in 2017. Full explanations and worked examples are in the study guide; test yourself with the practice set. The course hub is OCR A Level Physics, the printable list is the checklist, and the free 10-minute diagnostics show where to focus. X-rays and CAT scanning (6.5.1) are covered separately.

Tracers, 6.5.2(a)

Technetium-99m Fluorine-18
Emission Gamma only (about 140 keV) Positron (β⁺), then two gamma photons from annihilation
Half-life About 6 hours About 110 minutes
Used with Gamma camera PET scanner
Carried by Compound chosen for the organ Fluorodeoxyglucose, a glucose-like molecule

A tracer is a radioactive substance attached to a molecule that goes where the doctor wants to look. A good one emits gamma rays (they leave the body), and has a half-life long enough to use and short enough to limit dose. The “m” in Tc-99m means metastable.

Activity after time t: A = A0 × (1/2)^(t / half-life).

Gamma camera, 6.5.2(b) and (c)

Component Job
Collimator Lead plate with parallel holes; only photons travelling along the holes pass, so the image is not blurred
Scintillator Crystal that turns each absorbed gamma photon into a flash of light
Photomultiplier tubes Light releases electrons from a photocathode; dynodes multiply them into a pulse
Computer and display Compares pulses from tubes to find the position, adds a count, builds the image

Method for describing image formation, in steps:

  1. Tracer in the organ emits gamma photons in all directions.
  2. The collimator lets through only those travelling straight along its holes.
  3. A photon in the crystal gives a flash of light.
  4. The tubes nearest the flash give the biggest pulses.
  5. The computer works out the position and builds up the counts into an image.

Diagnosis: shows function, not just structure. Hot spots (high uptake) and cold spots (low uptake) show disease; images taken over time show how fast the organ takes up and clears the tracer.

PET scanner, 6.5.2(d) and (e)

  • Positron meets an electron and they annihilate: mass becomes two gamma photons.
  • Momentum conservation: the photons go in opposite directions.
  • Energy conservation: each carries the electron rest energy, mc² = 0.511 MeV (so 1.02 MeV in all).
  • A ring of detectors records photons arriving together. Each pair defines a line through the annihilation. The computer finds where many lines cross.
  • No collimator is used, so more of the radiation is detected than in a gamma camera.
  • Resolution is limited because the positron travels a short distance before it annihilates.
  • Diagnosis: fluorodeoxyglucose collects in tissue using much glucose, so tumours, the brain and the heart show brightly. Repeat scans show whether treatment has reduced activity.

Issues with an expensive scanner: purchase and running cost, a nearby supply of F-18 because of its short half-life, staff training, how many patients benefit, dose against benefit, and travel or waiting times without it.

Ultrasound, 6.5.3(a) to (c)

  • Longitudinal wave, frequency above 20 kHz. Scanners use a few MHz.
  • Higher frequency: shorter wavelength (λ = v/f), more detail, but more absorption, so less depth.
  • Piezoelectric effect: a p.d. across the crystal changes its shape, and deforming it gives a p.d.
  • A transducer is one crystal that emits pulses (alternating p.d. makes it vibrate) and receives echoes (vibration gives a p.d.).
  • A-scan: echo strength against time from one direction. Depth d = ct/2.
  • B-scan: many A-scans combined into a two-dimensional image; brightness shows echo strength.

Acoustic impedance and reflection, 6.5.3(d) to (f)

Quantity Equation Unit
Acoustic impedance Z = ρc kg m⁻² s⁻¹
Reflected fraction Ir/I0 = (Z2 − Z1)² / (Z2 + Z1)² none
Transmitted fraction 1 − Ir/I0 none

Method for a reflection question:

  1. Find Z for each medium, using Z = ρc if only density and speed are given.
  2. Subtract, add, then square the ratio.
  3. Convert to intensity or percent if asked.
  4. Comment: a big difference in Z means strong reflection and little transmission.

Gel: air has a tiny Z, skin a large one, so a bare air gap would reflect nearly everything. The gel removes air and has Z close to skin (impedance matching), so most ultrasound enters.

Doppler, 6.5.3(g)

Δf / f = 2v cos θ / c

  • θ is the angle between the beam and the blood flow; at 90° there is no shift.
  • The 2 arises because the moving blood acts as a receiver and then a source.
  • Rearranged: v = c Δf / (2 f cos θ).
  • Check the units: f and Δf in Hz (not MHz and kHz mixed).
  • Blood speed comes out in m s⁻¹ because c is in m s⁻¹.

Worked reminders

Decay. A Tc-99m source of 300 MBq is used 12 h later. That is 2 half-lives, so A = 300 × (1/2)² = 75 MBq.

Reflection. A beam of intensity 8.0 mW m⁻² crosses a boundary between tissues of Z = 1.38 × 10⁶ and 1.62 × 10⁶ kg m⁻² s⁻¹.

Ir/I0 = (0.24 × 10⁶ / 3.00 × 10⁶)² = 0.0064
Ir = 0.0064 × 8.0 = 0.051 mW m⁻²;  transmitted = 7.95 mW m⁻²

Doppler. A 6.0 MHz beam meets blood at 0.80 m s⁻¹, at 30° to the flow (c = 1540 m s⁻¹).

Δf = 2 × 6.0×10⁶ × 0.80 × cos 30° / 1540 = 5.4 × 10³ Hz

Image description. For any “describe how the image is formed” question, name each component in order and say what it does to the signal. For PET the order is: positron emitted, annihilation, two photons in opposite directions, detectors in a ring, coincidence, lines of response, computer.

Must-know distinctions

  • Gamma camera: single gamma photons, collimator needed. PET: pair of photons from annihilation, no collimator.
  • Tc-99m is a gamma emitter. F-18 is a positron emitter.
  • A-scan: a graph for one line. B-scan: a two-dimensional image.
  • Z is a property of the medium. Reflection depends on the difference between two Z values.
  • Gamma camera: shows where single photons came from by blocking the rest. PET: shows where pairs came from by timing.
  • Ultrasound intensity reflected and transmitted add up to the incident intensity at a boundary.

Quick self-test

  1. A Tc-99m source of 400 MBq is used 3.5 h later. Half-life 6.0 h. Find the activity.
  2. F-18 of 250 MBq is used 120 min later. Half-life 110 min. Find the activity.
  3. A gel has density 1030 kg m⁻³ and speed 1480 m s⁻¹. Find Z.
  4. Two media have Z = 1.50 × 10⁶ and 2.25 × 10⁶ kg m⁻² s⁻¹. Find Ir/I0.
  5. An echo arrives 52 μs after the pulse, with speed 1540 m s⁻¹. Find the depth.
  6. A 3.5 MHz beam at 40° gives a shift of 1.8 kHz (c = 1540 m s⁻¹). Find v.
  7. State the energy of one annihilation photon in MeV and the number of photons produced.
  8. Why does the gamma camera need a collimator?
  9. Why is F-18 made near the PET scanner?
  10. By what factor is the Doppler shift changed if the angle goes from 40° to 70°, other things equal?

Answers

  1. 400 × (1/2)^(3.5/6.0) = 267 MBq.
  2. 250 × (1/2)^(120/110) = 117 MBq.
  3. 1030 × 1480 = 1.52 × 10⁶ kg m⁻² s⁻¹.
  4. (0.75/3.75)² = 0.040.
  5. d = 1540 × 52 × 10⁻⁶ / 2 = 0.040 m (4.0 cm).
  6. v = 1540 × 1.8 × 10³ / (2 × 3.5 × 10⁶ × cos 40°) = 0.52 m s⁻¹.
  7. 0.511 MeV; two photons.
  8. It allows only photons travelling along the holes to reach the crystal, so each point in the image corresponds to a line through the patient.
  9. Its half-life of about 110 minutes means the activity falls quickly in transit.
  10. cos 70° / cos 40° = 0.45, so the shift falls to 45% of its value.

Where marks are usually lost

  • Giving the gel as a lubricant, with no mention of impedance or air.
  • Using d = ct rather than d = ct/2.
  • Dropping the 2 or cos θ in the Doppler equation, or mixing MHz and Hz.
  • Saying PET photons travel the same way: they go in opposite directions.
  • Saying the gamma camera detects positrons.
  • Writing the reflection fraction without squaring, or giving a negative answer.
  • Describing the collimator, scintillator and tubes without saying what each does to the signal.
  • Giving a one-sided answer when asked to discuss a scanner’s cost and benefit.

Official syllabus

OCR Advanced GCE in Physics A (H556), specification Version 3.0 (March 2026), for first assessment in 2017, published by OCR: sections 6.5.2 and 6.5.3.

Get free revision emails (optional)

Occasional emails with practice questions, worked explanations and links to free resources for the qualification and subjects you choose. No spam, and you can unsubscribe from any email. The free tools on this site never need an email.

Subjects (optional, up to 6)

Choose a qualification to see its subjects.

Related resources

Related articles

Studying this with a teacher

Working through Physics A LEVELS?

This page is free and stays free. If you would rather be taught it, Marlbridge runs Physics classes one-to-one, online in your own time zone. One-to-one only for this course: Rs 6,000 per class in Pakistan, with other currencies on the fees page. The first trial class is free. WhatsApp replies within an hour (8am–11pm Pakistan time, every day); email the same day.

OCR Physics teachers at Marlbridge