Revision Notes
AQA A-Level Physics (7408): Non-ionising imaging – Revision Notes
Condensed AQA A-level Physics 7408 notes on non-ionising imaging (3.10.4): impedance, A- and B-scans, endoscope fibre bundles, MR steps and a self-test.
- Subject
- Physics
- Level
- A LEVELS
- Topic
- Medical physics
- Author
- Marlbridge Academic Team
- Updated
- Reviewed by
- Hassan (what this means)
Aligned to AQA A Level Physics (7408), For first teaching 2015. Official specification .
Syllabus page (what it covers and how it is assessed): AQA A Level Physics.
Syllabus points this page covers
7408
- 3.10.4 Non-ionising imaging
Found an error? Report a correction.
Need help with this topic? Request a free trial class for A Level Physics (7408).
These notes condense section 3.10.4 Non-ionising imaging of the AQA AS and A-level Physics specification (7407/7408): ultrasound, fibre optics and endoscopy, and the MR scanner. It is part of the A-level only Medical physics option. For full explanations and worked examples read the non-ionising imaging study guide, then test yourself with the non-ionising imaging practice questions.
Equations
| Equation | Use | Notes |
|---|---|---|
| Z = ρc | Acoustic impedance | kg m⁻² s⁻¹ |
| I_r/I_i = ((Z₂ − Z₁)/(Z₂ + Z₁))² | Fraction reflected at a boundary | Normal incidence; transmitted = 1 − reflected |
| depth = ct/2 | Echo time to depth | t is the round-trip time |
| λ = c/f | Wavelength in tissue | Sets the finest detail |
| sin c = n₂/n₁ | Critical angle at core–cladding boundary | n₁ core > n₂ cladding |
Key terms
| Term | Meaning |
|---|---|
| Acoustic impedance | Product of density and speed of sound; decides reflection at a boundary |
| Attenuation | Loss of intensity with distance by absorption and scattering; greater at higher frequency |
| Piezoelectric effect | Crystal deforms when a p.d. is applied and produces a p.d. when deformed |
| Backing material | Damps the crystal so pulses are short |
| Coupling gel | Z close to skin; removes the air layer |
| A-scan | Echo amplitude against time along one line |
| B-scan | Swept probe; echo strength shown as dot brightness to form a 2D image |
| Coherent bundle | Fibres in the same relative positions at both ends; carries the image |
| Non-coherent bundle | Fibres randomly arranged; carries light in |
| Precession | Rotation of a proton’s spin axis about the magnetic field lines |
| Gradient coils | Vary the field across the patient to select the region scanned |
Ultrasound in one list
- Large Z difference → strong echo; gas and bone reflect most of the pulse, so they block imaging beyond them.
- Soft tissue boundaries reflect about 1%: most energy continues to deeper boundaries.
- Gel replaces air (air–skin reflects about 99.9%).
- Higher f: better resolution, more attenuation, less depth. Lower f: deeper, coarser.
- Transducer: alternating p.d. burst → vibration → pulse; damping → short pulse; echoes → p.d.
- Next pulse only after the deepest echo returns.
- Advantages: non-ionising, cheap, portable, real-time. Disadvantages: lower resolution, no imaging through bone or gas, operator-dependent.
Endoscope in one list
- Total internal reflection at the core–cladding interface keeps light in the core.
- Cladding: lower refractive index, protects the core surface, prevents leakage between fibres.
- Non-coherent bundle: light in. Coherent bundle: image out.
- Objective lens forms the image on the bundle end; eyepiece or camera views it.
- Thinner fibres → finer resolution.
- Channels for air, water and instruments → keyhole surgery: small incisions, faster recovery, lower infection risk.
MR scanner in six steps
- Superconducting magnet gives a strong field.
- Protons align with spins parallel and precess about the field lines.
- Gradient coils select the cross-section.
- Short RF pulses excite protons in successive small regions (change of spin state).
- Protons de-excite and emit RF signals.
- Signals detected and processed by computer into an image.
Not examined: how the fields are produced, relaxation times.
Comparison table (resolution, convenience, safety)
| Method | Resolution | Convenience | Safety |
|---|---|---|---|
| Ultrasound | Lower | Cheap, portable, live | Non-ionising |
| Endoscopy | Direct view of surfaces | Invasive but minor | Non-ionising; small infection risk |
| MR | High, excellent soft tissue | Expensive, slow, noisy, confined | Non-ionising; metal implants dangerous |
Exam traps
- Depth = ct, missing the 2.
- Squaring only the numerator in the reflection formula.
- Saying gel has “low impedance”; it has impedance close to skin.
- Swapping the roles of the coherent and non-coherent bundles.
- Saying MR images electrons or uses X-rays.
- Saying lower frequency gives better resolution.
Self-test
- Calculate Z for tissue with density 1040 kg m⁻³ and speed of sound 1570 m s⁻¹.
- Calculate the fraction reflected at a fat (Z = 1.38 × 10⁶) to blood (Z = 1.66 × 10⁶) boundary.
- An echo returns after 130 µs (c = 1540 m s⁻¹). How deep is the boundary?
- Calculate the wavelength of 5.0 MHz ultrasound in tissue (c = 1540 m s⁻¹).
- Why is backing material placed behind the piezoelectric crystal?
- State the difference between an A-scan and a B-scan.
- Why can ultrasound not image the lungs?
- Calculate the critical angle for a core of n = 1.52 and cladding of n = 1.46.
- Why must the image-carrying bundle be coherent?
- What is the role of the gradient coils in an MR scanner?
- What do protons emit as they de-excite?
- Give one safety disadvantage of MR scanning.
Answers
- 1040 × 1570 = 1.63 × 10⁶ kg m⁻² s⁻¹.
- (0.28/3.04)² = 8.5 × 10⁻³ (0.85%).
- 1540 × 130 × 10⁻⁶/2 = 0.100 m.
- 1540/5.0 × 10⁶ = 3.1 × 10⁻⁴ m (0.31 mm).
- To damp the vibration so each pulse is short, letting close boundaries be distinguished.
- A-scan: amplitude against time along one line, probe fixed. B-scan: probe swept, echoes shown as dots of varying brightness forming a 2D image.
- The lungs contain air, whose impedance is very different from tissue, so almost all the ultrasound is reflected.
- sin c = 1.46/1.52, so c = 73.8°.
- Each fibre must arrive at the same relative position at the viewing end, so the pattern of light (the image) is preserved.
- They make the field vary across the patient, selecting the cross-section and the small regions excited in turn.
- Radio-frequency (RF) signals.
- The strong field can move or heat metal implants or affect pacemakers (also: confined space).
Frequently asked questions
Why does higher frequency ultrasound give better images of shallow structures? Higher frequency means a shorter wavelength, so finer detail is resolved. Attenuation also rises with frequency, so high frequencies are used only where the structure is close to the probe.
What is the difference between coherent and non-coherent fibre bundles? A coherent bundle keeps each fibre in the same relative position at both ends, so it can carry an image. A non-coherent bundle has randomly arranged fibres and only carries light into the body.
Official syllabus
AQA AS and A-level Physics specification (7407/7408), AS and A-level exams June 2016 onwards: section 3.10.4 Non-ionising imaging (A-level only).
Related resources
-
Study guides
AQA A-Level Physics (7408): Physics of the ear
AQA A-level Physics 7408 section 3.10.2 in depth: how the ear transmits sound, decibel and dBA calculations, equal loudness curves and hearing loss.
Physics · AQA · A Level
Read guide -
Practice questions
AQA A-Level Physics (7408): Physics of the ear – Practice Questions
Original AQA A-level Physics 7408 practice questions on the ear: transmission, pressure gain, decibel calculations, equal loudness curves and hearing loss.
Physics · AQA · A Level
Try questions -
Revision notes
AQA A-Level Physics (7408): Physics of the ear – Revision Notes
Condensed AQA A-level Physics 7408 notes on the ear (3.10.2): transmission chain, decibel rules, equal loudness curves, dBA, hearing loss and a self-test.
Physics · AQA · A Level
Revise topic
Related articles
Get free revision emails (optional)
Occasional emails with practice questions, worked explanations and links to free resources for the qualification and subjects you choose. No spam, and you can unsubscribe from any email. The free tools on this site never need an email.