AQA A-Level Physics (7408): Non-ionising imaging – Revision Notes

Condensed AQA A-level Physics 7408 notes on non-ionising imaging (3.10.4): impedance, A- and B-scans, endoscope fibre bundles, MR steps and a self-test.

Subject
Physics
Level
A LEVELS
Topic
Medical physics
Updated

Aligned to AQA A Level Physics (7408), For first teaching 2015. Official specification .

Syllabus page (what it covers and how it is assessed): AQA A Level Physics.

Syllabus points this page covers

7408

  • 3.10.4 Non-ionising imaging

Found an error? Report a correction.

Need help with this topic? Request a free trial class for A Level Physics (7408).

These notes condense section 3.10.4 Non-ionising imaging of the AQA AS and A-level Physics specification (7407/7408): ultrasound, fibre optics and endoscopy, and the MR scanner. It is part of the A-level only Medical physics option. For full explanations and worked examples read the non-ionising imaging study guide, then test yourself with the non-ionising imaging practice questions.

Equations

Equation Use Notes
Z = ρc Acoustic impedance kg m⁻² s⁻¹
I_r/I_i = ((Z₂ − Z₁)/(Z₂ + Z₁))² Fraction reflected at a boundary Normal incidence; transmitted = 1 − reflected
depth = ct/2 Echo time to depth t is the round-trip time
λ = c/f Wavelength in tissue Sets the finest detail
sin c = n₂/n₁ Critical angle at core–cladding boundary n₁ core > n₂ cladding

Key terms

Term Meaning
Acoustic impedance Product of density and speed of sound; decides reflection at a boundary
Attenuation Loss of intensity with distance by absorption and scattering; greater at higher frequency
Piezoelectric effect Crystal deforms when a p.d. is applied and produces a p.d. when deformed
Backing material Damps the crystal so pulses are short
Coupling gel Z close to skin; removes the air layer
A-scan Echo amplitude against time along one line
B-scan Swept probe; echo strength shown as dot brightness to form a 2D image
Coherent bundle Fibres in the same relative positions at both ends; carries the image
Non-coherent bundle Fibres randomly arranged; carries light in
Precession Rotation of a proton’s spin axis about the magnetic field lines
Gradient coils Vary the field across the patient to select the region scanned

Ultrasound in one list

  • Large Z difference → strong echo; gas and bone reflect most of the pulse, so they block imaging beyond them.
  • Soft tissue boundaries reflect about 1%: most energy continues to deeper boundaries.
  • Gel replaces air (air–skin reflects about 99.9%).
  • Higher f: better resolution, more attenuation, less depth. Lower f: deeper, coarser.
  • Transducer: alternating p.d. burst → vibration → pulse; damping → short pulse; echoes → p.d.
  • Next pulse only after the deepest echo returns.
  • Advantages: non-ionising, cheap, portable, real-time. Disadvantages: lower resolution, no imaging through bone or gas, operator-dependent.

Endoscope in one list

  • Total internal reflection at the core–cladding interface keeps light in the core.
  • Cladding: lower refractive index, protects the core surface, prevents leakage between fibres.
  • Non-coherent bundle: light in. Coherent bundle: image out.
  • Objective lens forms the image on the bundle end; eyepiece or camera views it.
  • Thinner fibres → finer resolution.
  • Channels for air, water and instruments → keyhole surgery: small incisions, faster recovery, lower infection risk.

MR scanner in six steps

  1. Superconducting magnet gives a strong field.
  2. Protons align with spins parallel and precess about the field lines.
  3. Gradient coils select the cross-section.
  4. Short RF pulses excite protons in successive small regions (change of spin state).
  5. Protons de-excite and emit RF signals.
  6. Signals detected and processed by computer into an image.

Not examined: how the fields are produced, relaxation times.

Comparison table (resolution, convenience, safety)

Method Resolution Convenience Safety
Ultrasound Lower Cheap, portable, live Non-ionising
Endoscopy Direct view of surfaces Invasive but minor Non-ionising; small infection risk
MR High, excellent soft tissue Expensive, slow, noisy, confined Non-ionising; metal implants dangerous

Exam traps

  • Depth = ct, missing the 2.
  • Squaring only the numerator in the reflection formula.
  • Saying gel has “low impedance”; it has impedance close to skin.
  • Swapping the roles of the coherent and non-coherent bundles.
  • Saying MR images electrons or uses X-rays.
  • Saying lower frequency gives better resolution.

Self-test

  1. Calculate Z for tissue with density 1040 kg m⁻³ and speed of sound 1570 m s⁻¹.
  2. Calculate the fraction reflected at a fat (Z = 1.38 × 10⁶) to blood (Z = 1.66 × 10⁶) boundary.
  3. An echo returns after 130 µs (c = 1540 m s⁻¹). How deep is the boundary?
  4. Calculate the wavelength of 5.0 MHz ultrasound in tissue (c = 1540 m s⁻¹).
  5. Why is backing material placed behind the piezoelectric crystal?
  6. State the difference between an A-scan and a B-scan.
  7. Why can ultrasound not image the lungs?
  8. Calculate the critical angle for a core of n = 1.52 and cladding of n = 1.46.
  9. Why must the image-carrying bundle be coherent?
  10. What is the role of the gradient coils in an MR scanner?
  11. What do protons emit as they de-excite?
  12. Give one safety disadvantage of MR scanning.

Answers

  1. 1040 × 1570 = 1.63 × 10⁶ kg m⁻² s⁻¹.
  2. (0.28/3.04)² = 8.5 × 10⁻³ (0.85%).
  3. 1540 × 130 × 10⁻⁶/2 = 0.100 m.
  4. 1540/5.0 × 10⁶ = 3.1 × 10⁻⁴ m (0.31 mm).
  5. To damp the vibration so each pulse is short, letting close boundaries be distinguished.
  6. A-scan: amplitude against time along one line, probe fixed. B-scan: probe swept, echoes shown as dots of varying brightness forming a 2D image.
  7. The lungs contain air, whose impedance is very different from tissue, so almost all the ultrasound is reflected.
  8. sin c = 1.46/1.52, so c = 73.8°.
  9. Each fibre must arrive at the same relative position at the viewing end, so the pattern of light (the image) is preserved.
  10. They make the field vary across the patient, selecting the cross-section and the small regions excited in turn.
  11. Radio-frequency (RF) signals.
  12. The strong field can move or heat metal implants or affect pacemakers (also: confined space).

Frequently asked questions

Why does higher frequency ultrasound give better images of shallow structures? Higher frequency means a shorter wavelength, so finer detail is resolved. Attenuation also rises with frequency, so high frequencies are used only where the structure is close to the probe.

What is the difference between coherent and non-coherent fibre bundles? A coherent bundle keeps each fibre in the same relative position at both ends, so it can carry an image. A non-coherent bundle has randomly arranged fibres and only carries light into the body.