AQA A-Level Physics (7408): Non-ionising imaging
AQA A-level Physics 7408 section 3.10.4 in depth: ultrasound impedance and scans, piezoelectric probes, fibre-optic endoscopes and the MR scanner.
- Subject
- Physics
- Level
- A LEVELS
- Topic
- Medical physics
- Author
- Marlbridge Academic Team
- Updated
- Reviewed by
- Hassan (what this means)
Aligned to AQA A Level Physics (7408), For first teaching 2015. Official specification .
Syllabus page (what it covers and how it is assessed): AQA A Level Physics.
Syllabus points this page covers
7408
- 3.10.4 Non-ionising imaging
Found an error? Report a correction.
Need help with this topic? Request a free trial class for A Level Physics (7408).
This guide goes deeper into section 3.10.4 Non-ionising imaging of the AQA AS and A-level Physics specification (7407/7408), part of the A-level only Medical physics option assessed in Paper 3 Section B. It has three content statements: 3.10.4.1 ultrasound imaging, 3.10.4.2 fibre optics and endoscopy, and 3.10.4.3 the magnetic resonance (MR) scanner. None of these uses ionising radiation, which is their shared safety advantage over X-rays and radionuclides.
The Medical physics study guide gives the option overview. For recall use the non-ionising imaging revision notes, and test yourself with the non-ionising imaging practice questions. Total internal reflection is first met in Waves.
3.10.4.1 Ultrasound imaging
Acoustic impedance and reflection
Ultrasound is sound above about 20 kHz; medical scanners use frequencies of a few megahertz. When a pulse meets a boundary between two tissues, part is reflected and part transmitted. The deciding property is the acoustic impedance:
Z = ρc (kg m⁻² s⁻¹), where ρ is density and c the speed of sound in the material.
The fraction of intensity reflected at normal incidence is
I_r/I_i = ((Z₂ − Z₁)/(Z₂ + Z₁))²
and the rest is transmitted. A large difference in Z gives a strong echo; similar values of Z give a weak echo and good transmission.
Worked example 1. Use these illustrative impedances (all × 10⁶ kg m⁻² s⁻¹): coupling gel 1.60, skin 1.70, fat 1.38, muscle 1.70, bone 7.80; air 430 kg m⁻² s⁻¹.
gel → skin: ((1.70 − 1.60)/(3.30))² = 9.2 × 10⁻⁴ (almost all transmitted)
fat → muscle: ((1.70 − 1.38)/(3.08))² = 0.011 (weak but useful echo)
muscle → bone: ((7.80 − 1.70)/(9.50))² = 0.41 (strong echo)
air → skin: ((1.70 × 10⁶ − 430)/(1.70 × 10⁶ + 430))² = 0.999
Three conclusions follow, and each is a common exam point:
- Without gel, a thin air layer between probe and skin reflects 99.9% of the pulse, so coupling gel, with Z close to that of skin, is essential.
- Soft-tissue boundaries reflect about 1%, so most of the pulse continues and later boundaries can still be imaged.
- Bone reflects a large fraction and gas reflects almost everything, so ultrasound cannot image through bone or air-filled organs such as the lungs.
Attenuation
As the pulse travels, its intensity falls because energy is absorbed and scattered by tissue. Attenuation increases with frequency. This creates the central trade-off of ultrasound:
- Higher frequency → shorter wavelength → better resolution (finer detail), but more attenuation, so less depth.
- Lower frequency → deeper penetration but poorer resolution.
At 3.5 MHz in soft tissue (c = 1540 m s⁻¹), λ = 1540/3.5 × 10⁶ = 0.44 mm; at 10 MHz, λ = 0.15 mm. Deep abdominal scans use the lower frequency; shallow structures such as the eye use the higher one. Scanners also amplify later echoes more than early ones to compensate for attenuation.
Piezoelectric transducers
A piezoelectric crystal changes shape when a p.d. is applied across it, and produces a p.d. when it is deformed. One crystal does both jobs:
- Generation: a short burst of alternating p.d. at the crystal’s resonant frequency makes it vibrate and emit an ultrasound pulse.
- Damping: backing material behind the crystal stops the vibration quickly, so the pulse is short. Short pulses let echoes from close boundaries be distinguished.
- Detection: between pulses the crystal acts as a receiver; returning echoes make it vibrate and produce a small p.d. that is amplified.
The probe must wait for the echoes from the deepest boundary before sending the next pulse. For a depth of 0.20 m the round trip takes 2 × 0.20/1540 = 260 µs, so at most about 3800 pulses can be sent each second.
A-scans and B-scans
- A-scan (amplitude scan): the probe is held still, and echo amplitude is displayed against time. Depth = ct/2, where t is the echo time. Used to measure distances along one line, such as the depth of structures in the eye.
- B-scan (brightness scan): the probe is moved or swept, often using an array of crystals. Each echo is shown as a dot whose brightness depends on the echo strength, at a position set by its depth and the probe direction. The dots build a two-dimensional image, used in fetal scanning and for imaging organs.
Worked example 2. An A-scan through a section of tissue shows echoes from its front and back surfaces at 40 µs and 95 µs. Find the depth of each surface and the thickness (c = 1540 m s⁻¹).
front: 1540 × 40 × 10⁻⁶/2 = 0.0308 m = 30.8 mm
back: 1540 × 95 × 10⁻⁶/2 = 0.0732 m = 73.2 mm
thickness = 42.4 mm
Comparing ultrasound with alternatives
| Advantages | Disadvantages |
|---|---|
| Non-ionising, so safe for fetuses and repeat scans | Lower resolution than CT or MR |
| Cheap and portable | Cannot pass through bone or gas |
| Real-time images of movement | Image quality depends on the operator and on the patient’s build |
3.10.4.2 Fibre optics and endoscopy
An optical fibre has a core of higher refractive index surrounded by a cladding of lower refractive index. Light hitting the core–cladding boundary at more than the critical angle is totally internally reflected, so it travels along the fibre even when it bends. The cladding protects the core surface and stops light leaking between touching fibres.
Worked example 3. Core n = 1.50, cladding n = 1.47: sin c = 1.47/1.50 = 0.980, so c = 78.5°. Rays must strike the boundary at more than 78.5° to the normal; this limits how sharply the fibre can be bent before light escapes.
A flexible endoscope contains two fibre bundles:
- A non-coherent bundle carries light from an external source into the body. The fibres can be arranged randomly because only light, not an image, is carried; it is cheaper to make.
- A coherent bundle carries the image out. The fibres keep the same relative positions at both ends, so each fibre carries one small part of the image to the matching place at the other end. An objective lens at the distal end forms an image on the end of the bundle; an eyepiece or camera views the image at the other end.
Resolution depends on fibre diameter: thinner fibres give more picture elements and finer detail. The endoscope tube also has channels for water, air and surgical instruments. Uses include viewing the stomach, airways or joints and keyhole surgery, which needs only small incisions, so patients recover faster with less risk of infection than after open surgery.
3.10.4.3 The MR scanner
The specification lists the principles to know:
- The patient lies in the strong, uniform field of a superconducting magnet.
- Hydrogen nuclei (protons) in the body have spin; they align with spins parallel to the field and precess about the field lines.
- Gradient field coils add a field that varies across the patient, so only one cross-section (and, within it, one small region at a time) meets the conditions for excitation.
- Short radio-frequency (RF) pulses excite protons in successive small regions, changing their spin state.
- As the excited protons de-excite, they emit RF signals.
- The RF signals are detected and processed by a computer to build a visual image of the cross-section.
Different tissues contain different amounts of hydrogen, mostly in water and fat, so soft tissues are distinguished very well. Production of the fields and relaxation times are not examined.
Comparing MR: no ionising radiation and excellent soft-tissue detail with high resolution in any plane; but the scanner is expensive, scans are slow and noisy, the patient must stay still in a confined tube, and metal implants or pacemakers can be dangerous in the strong field.
Model answer
Question: Explain why a gel is applied between an ultrasound probe and the skin. [2]
Answer: Without gel there is air between probe and skin; the acoustic impedance of air is very much smaller than that of skin, so almost all the ultrasound is reflected at the boundary [1]. The gel has an acoustic impedance close to that of skin, so almost all the ultrasound is transmitted into the body [1].
Common mistakes
- Forgetting the 2 in depth = ct/2.
- Saying the coherent bundle carries light in; it carries the image out.
- Saying ultrasound is used for lungs or bones.
- Describing MR in terms of X-rays or ionisation.
- Claiming higher frequency penetrates further.
Exam technique
- Calculate reflected fractions to 2 or 3 significant figures, then comment on what the number means for the image.
- Explain coherent bundles with “same relative positions at each end”.
- Describe MR as a sequence of the six steps above.
- Compare: the specification limits comparisons to resolution, convenience and safety, so organise answers under those headings.
Next, try the non-ionising imaging practice questions.
Official syllabus
AQA AS and A-level Physics specification (7407/7408), AS and A-level exams June 2016 onwards: section 3.10.4 Non-ionising imaging (A-level only), content statements 3.10.4.1 to 3.10.4.3.
Related resources
-
Practice questions
AQA A-Level Physics (7408): Non-ionising imaging – Practice Questions
Original AQA A-level Physics 7408 practice on non-ionising imaging: acoustic impedance, echo depths, ultrasound frequency choice, endoscopes and MR scanning.
Physics · AQA · A Level
Try questions -
Revision notes
AQA A-Level Physics (7408): Non-ionising imaging – Revision Notes
Condensed AQA A-level Physics 7408 notes on non-ionising imaging (3.10.4): impedance, A- and B-scans, endoscope fibre bundles, MR steps and a self-test.
Physics · AQA · A Level
Revise topic -
Study guides
AQA A-Level Physics (7408): Physics of the ear
AQA A-level Physics 7408 section 3.10.2 in depth: how the ear transmits sound, decibel and dBA calculations, equal loudness curves and hearing loss.
Physics · AQA · A Level
Read guide
Related articles
Get free revision emails (optional)
Occasional emails with practice questions, worked explanations and links to free resources for the qualification and subjects you choose. No spam, and you can unsubscribe from any email. The free tools on this site never need an email.