Cambridge A Level Information Technology (9626): Monitoring and measurement technologies
Cambridge 9626 IT section 3.1 study guide: the named sensors, environmental and patient monitoring, and one-point, two-point and multipoint calibration.
- Subject
- ICT
- Level
- AS LEVEL
- Topic
- Topic 3 – Monitoring and control
- Author
- Marlbridge Academic Team
- Updated
- Reviewed by
- Harris Khan (what this means)
Aligned to Cambridge A Level Information Technology (9626), 2025-2027. Official specification .
Syllabus page (what it covers and how it is assessed): Cambridge A Level Information Technology.
Syllabus points this page covers
9626 (AS Level)
- 3.1 Monitoring and measurement technologies
Found an error? Report a correction.
Need help with this topic? Request a free trial class for A Level Information Technology (9626).
This guide teaches section 3.1 Monitoring and measurement technologies of Topic 3 Monitoring and control in the Cambridge International AS & A Level Information Technology 9626 syllabus for 2025, 2026 and 2027. It is AS Level content, examined in Paper 1 Theory. The Monitoring and control overview separates monitoring from control; this page covers every sensor and use in 3.1 and works through calibration with numbers. Use it with the monitoring and measurement revision notes and the monitoring and measurement practice questions.
What section 3.1 covers
| Syllabus point | Content |
|---|---|
| Sensors | Light/UV, temperature, pressure, humidity, pH, gas (oxygen, carbon dioxide, carbon monoxide, oxides of nitrogen), sound, infrared, touch, (electro)magnetic field, proximity |
| Environmental monitoring | Water pollution; weather stations and their sensors; ambient temperature; atmospheric pressure; absolute and relative humidity; sunlight |
| Patient monitoring | Monitoring patients |
| Calibration | Why it matters; when to use one-point, two-point and multipoint calibration; how readings from sensors such as temperature sensors are calibrated |
How a monitoring system works
- Sensors measure a physical quantity continuously or at set intervals.
- Most sensors produce an analogue signal, so an analogue-to-digital converter turns it into digital data the computer can process.
- The computer stores the readings, often with the time of each one, and displays them as tables or graphs.
- It compares readings with preset limits and warns a person, for example with an alarm, if a value is out of range.
A monitoring system does not change the environment itself: a person decides what to do. When the computer acts automatically, for example opening a valve, it has become a control system (section 3.2).
Two settings are chosen for every monitoring task: the sampling interval (time between readings) and the period of monitoring (how long readings are taken for). A patient’s heart rate might be read every second for days; a river’s pH might be read every 15 minutes for a year.
The sensors named in 3.1
| Sensor | What it measures | Example in monitoring |
|---|---|---|
| Light/UV | Light intensity, or ultraviolet levels | Sunlight at a weather station; UV levels for health warnings |
| Temperature | How hot or cold something is | Ambient air temperature; a patient’s body temperature; river temperature |
| Pressure | Force on a surface | Atmospheric pressure; blood pressure |
| Humidity | Water vapour in the air | Weather stations; museum display cases |
| pH | Acidity or alkalinity of a liquid | Water pollution in rivers and lakes |
| Gas: oxygen | Oxygen level in air or dissolved in water | Fish health in a river; air in a mine |
| Gas: carbon dioxide | CO2 level | Air quality in classrooms and offices |
| Gas: carbon monoxide | CO, a poisonous gas | Homes, car parks, workshops |
| Gas: oxides of nitrogen | NOx, mainly from vehicle exhausts | Roadside air pollution |
| Sound | Sound level | Noise near airports and building sites |
| Infrared | Infrared radiation, such as body heat | Detecting people; measuring temperature without contact |
| Touch | Contact or pressure from a finger or object | Detecting whether a sensor probe is in contact with a surface |
| (Electro)magnetic field | Strength of a magnetic field | Detecting passing vehicles or metal objects; checking field levels near equipment |
| Proximity | How close an object is, without contact | Detecting how near a person or object is to equipment |
Environmental monitoring
Water pollution
Sensors are placed in rivers, lakes or outflow pipes, often in a floating or fixed unit that sends data wirelessly.
- pH sensors detect acid or alkaline pollution, such as chemicals from a factory.
- Temperature sensors detect warm water discharged from power stations or factories, which holds less dissolved oxygen.
- Oxygen sensors measure dissolved oxygen; low levels harm fish and can signal sewage or fertiliser pollution.
The system logs readings day and night, so a short pollution event that a person taking occasional samples would miss is still recorded, with its exact time.
Weather stations
A weather station combines several sensors and logs their readings automatically, often in remote places.
- Temperature sensors measure ambient temperature, the temperature of the surrounding air. They are shaded and ventilated so direct sunlight does not heat them.
- Pressure sensors measure atmospheric pressure. Falling pressure often signals unsettled weather; rising pressure, settled weather.
- Humidity sensors measure water vapour. Absolute humidity is the actual mass of water vapour in a given volume of air, for example in grams per cubic metre. Relative humidity is the amount of water vapour compared with the maximum the air could hold at that temperature, given as a percentage. Warm air can hold more water, so the same absolute humidity gives a lower relative humidity on a warmer day.
- Light sensors measure sunlight, recording how strong it is and for how many hours the sun shines.
Readings are sent to a central computer that combines data from many stations to produce forecasts and long-term records.
Monitoring patients
In hospital wards and intensive care, sensors attached to the patient feed a bedside monitor:
- heart rate (pulse) and the heart’s electrical activity
- blood pressure, using a pressure sensor
- body temperature, using a temperature sensor
- blood oxygen level, using a sensor that shines light through a finger or ear lobe
- breathing rate
The monitor displays the values and graphs, stores them, and sounds an alarm when any value moves outside limits set for that patient.
Advantages: readings are continuous, day and night; alarms respond instantly; nurses are freed from manual checks for other care; readings are not affected by human error or tiredness; one nurse can watch several patients from a central station.
Disadvantages: equipment is expensive; false alarms, for example when a patient moves and a sensor slips, can cause alarm fatigue; staff may rely on the machine and miss signs it does not measure; faulty or badly calibrated sensors give misleading readings.
Calibration
Calibration is checking a sensor’s readings against known, accurate reference values and adjusting the system so that its readings match. It matters because:
- sensors drift over time through age, wear and temperature changes
- two sensors of the same model can give slightly different readings
- decisions rest on the readings: a patient’s medication, a pollution warning, a weather forecast
One-point calibration
The sensor is checked against one reference value, and the difference (the offset) is added to or subtracted from every reading. It suits a sensor whose readings are all wrong by the same amount, or one used only close to a single value.
Worked example 1. A room temperature sensor reads 21.6 °C when an accurate reference thermometer shows 20.0 °C. The offset is −1.6 °C. A later raw reading of 25.3 °C is corrected to 25.3 − 1.6 = 23.7 °C.
Two-point calibration
The sensor is checked at two reference values, usually near the bottom and top of its working range. This corrects both the offset and the slope (whether the sensor over- or under-reads more as the value rises). It suits sensors that respond in a straight line but may have both kinds of error.
A temperature sensor can be calibrated in a mixture of ice and water (0 °C) and in boiling water (100 °C at standard atmospheric pressure).
Worked example 2. A sensor reads 1.5 in the ice bath and 98.0 in boiling water. Its raw range is 98.0 − 1.5 = 96.5 for a true range of 100.
Corrected value = (raw reading − 1.5) × 100 ÷ 96.5
A raw reading of 37.0 gives (37.0 − 1.5) × 100 ÷ 96.5 = 36.8 °C (1 decimal place). Notice that a one-point correction at 0 °C alone would give 35.5 °C, which is wrong by more than a degree for a body temperature.
Multipoint calibration
The sensor is checked at three or more reference values across its range. Readings between two calibration points are corrected using those two points. It suits sensors whose response is not a straight line and situations where high accuracy is needed across the whole range.
Worked example 3. A sensor is checked at five points:
| True value (°C) | 0 | 25 | 50 | 75 | 100 |
|---|---|---|---|---|---|
| Raw reading | 0.8 | 26.0 | 52.4 | 79.6 | 103.0 |
A raw reading of 62.0 lies between 52.4 (true 50) and 79.6 (true 75). Corrected = 50 + (62.0 − 52.4) × 25 ÷ (79.6 − 52.4) = 58.8 °C. A single straight line through the end points would not fit the middle readings this well.
Choosing a method
| Method | Use when |
|---|---|
| One-point | Readings are off by a constant amount, or the sensor is used near one value |
| Two-point | The response is a straight line, used across a wide range |
| Multipoint | The response is not linear, or high accuracy is needed across the full range |
Common mistakes
- Describing an automatic response in a monitoring question: monitoring warns a person; control acts.
- Mixing up absolute and relative humidity.
- Saying calibration “makes the sensor more sensitive”. It makes readings accurate.
- Choosing one-point calibration for a sensor with a slope error.
- Naming a sensor without saying what it measures in the scenario.
Exam technique
- Describe a monitoring system: sensors → analogue-to-digital conversion → store, display, compare → warn.
- Explain a sensor’s use: what it measures and why that matters in the scenario.
- Calibration questions: name the method, say when it is suitable, and show the working when figures are given.
Official syllabus
Cambridge International AS & A Level Information Technology 9626 syllabus for examination in 2025, 2026 and 2027, Cambridge International Education. Topic 3 Monitoring and control, section 3.1 Monitoring and measurement technologies.
Related resources
-
Practice questions
Cambridge A Level Information Technology (9626): Monitoring and measurement technologies – Practice Questions
Twelve original Cambridge 9626 IT section 3.1 questions with answers: sensors, humidity, water and patient monitoring, and calibration calculations.
ICT · Cambridge · AS Level
Try questions -
Revision notes
Cambridge A Level Information Technology (9626): Monitoring and measurement technologies – Revision Notes
Revision notes for Cambridge 9626 IT section 3.1: sensors, weather and water monitoring, patient monitoring, humidity and calibration methods, with a self-test.
ICT · Cambridge · AS Level
Revise topic -
Study guides
Cambridge A Level Information Technology (9626): Control technologies
Cambridge 9626 IT section 3.2 study guide: control sensors, the nine actuator types, named control systems, trade-offs and control algorithms traced.
ICT · Cambridge · AS Level
Read guide
Get free revision emails (optional)
Occasional emails with practice questions, worked explanations and links to free resources for the qualification and subjects you choose. No spam, and you can unsubscribe from any email. The free tools on this site never need an email.