Syllabus points this page covers

9702 (A Level)

  • 24.1 Production and use of ultrasound

Found an error? Report a correction.

Need help with this topic? Request a free trial class for A Level Physics (9702).

Subtopic 24.1 of Cambridge International AS & A Level Physics 9702 (2025–2027) covers the piezo-electric effect, generation and detection of ultrasound, pulse-echo diagnosis, specific acoustic impedance, the intensity reflection coefficient and attenuation. These notes are condensed for recall. For full explanations and worked examples, read the Ultrasound study guide; to apply the ideas, try the Ultrasound practice questions.

Equations

Z = ρc                                 unit kg m⁻² s⁻¹
I_R / I₀ = (Z₁ − Z₂)² / (Z₁ + Z₂)²     (on the data sheet)
I = I₀ e^(−μx)                          (recall)
depth of boundary d = ct / 2

Piezo-electric effect: both directions

Cause Effect Used for
p.d. applied across crystal Crystal changes shape Generating ultrasound
Crystal’s shape changed e.m.f. generated across crystal Detecting ultrasound

Generation and detection sequence

Generating:

  1. Alternating p.d. applied across the crystal.
  2. Crystal expands and contracts at that frequency.
  3. Vibrating face produces longitudinal ultrasound waves.
  4. Largest amplitude at resonance: applied frequency = natural frequency of crystal (set by its thickness).

Detecting:

  1. Returning wave produces pressure changes at the crystal.
  2. Crystal is compressed and stretched.
  3. Alternating e.m.f. generated, then amplified.

Pulses are used so the same transducer can switch to receiving, and so the echo time can be measured. Damping material makes each pulse short.

Diagnostic information from echoes

Measurement Tells you
Time delay of echo Depth of boundary, d = ct/2
Intensity of echo How different the two tissues are (Z mismatch), after attenuation
Many echoes from a scan Positions of boundaries, built into an image

Reflection: key comparisons

Boundary Z mismatch Fraction reflected Consequence
Air and skin Huge About 0.999 Coupling gel essential
Gel and skin Small About 0.001 Almost all enters the body
Soft tissue and bone Large About 0.4 Strong echo; little reaches tissue behind bone
Fat and muscle Small About 0.015 Weak echo, needs amplification

Fraction transmitted = 1 − I_R/I₀.

Attenuation

  • Intensity falls exponentially with distance because of absorption and scattering.
  • μ is the attenuation coefficient (m⁻¹).
  • For an echo from depth x, the path is 2x: I/I₀ = e^(−2μx).
  • Echo fraction = (reflection coefficient) × (attenuation factor): multiply, do not add.
  • Thickness that halves the intensity: x = ln 2 / μ (for μ = 23 m⁻¹, about 3.0 cm).

Calculation chain for an echo

  1. Z = ρc for each tissue (if not given).
  2. Reflection coefficient at the boundary from the two Z values.
  3. Attenuation factor e^(−μ × 2x) for the journey there and back.
  4. Echo intensity = I₀ × reflection coefficient × attenuation factor.
  5. Depth from the echo time: d = ct/2.

Key terms

Term Meaning
Ultrasound Longitudinal sound wave with frequency above about 20 kHz
Transducer Device converting electrical energy to ultrasound and back
Specific acoustic impedance Density of a medium × speed of sound in it
Intensity reflection coefficient Fraction of incident intensity reflected at a boundary
Attenuation Decrease in intensity as the wave passes through matter
Coupling gel Gel with Z close to skin, removing air between transducer and skin
Impedance matching Using media of similar Z so that most intensity is transmitted

Exam traps

  • Saying a current, rather than a p.d., changes the crystal’s shape.
  • Describing generation only, and forgetting detection by the same crystal.
  • d = ct instead of d = ct/2.
  • Attenuating over x instead of 2x for an echo.
  • “Gel acts as a lubricant”: wrong reason.
  • Leaving the 10⁶ in one impedance but not the other.
  • Writing Z in kg m⁻³ (density units).

Self-test

  1. What happens to a piezo-electric crystal when a p.d. is applied across it?
  2. What happens when its shape is changed by a sound wave?
  3. Why must the applied p.d. be alternating to produce ultrasound?
  4. Why is ultrasound sent in short pulses?
  5. Define specific acoustic impedance.
  6. Calculate Z for a tissue of density 1040 kg m⁻³ in which sound travels at 1570 m s⁻¹.
  7. An echo returns 32 μs after the pulse. Sound travels at 1540 m s⁻¹. How deep is the boundary?
  8. Explain why a coupling gel is used.
  9. Calculate the reflection coefficient between media of Z = 1.34 × 10⁶ and 1.71 × 10⁶ kg m⁻² s⁻¹.
  10. Ultrasound passes 5.0 cm into tissue with μ = 23 m⁻¹. What fraction of the intensity remains?
  11. Why is it difficult to image organs lying behind bone?
  12. Two echoes from the front and back of an organ arrive 32 μs apart. With c = 1540 m s⁻¹, how thick is the organ?

Answers: 1. It changes shape (becomes thicker or thinner). 2. An e.m.f. is generated across it. 3. The crystal must expand and contract repeatedly to produce a wave of that frequency. 4. So the crystal can stop and receive echoes, and so the echo time can be measured. 5. The product of the density of a medium and the speed of sound in it, Z = ρc. 6. 1040 × 1570 = 1.63 × 10⁶ kg m⁻² s⁻¹. 7. d = 1540 × 32 × 10⁻⁶ / 2 = 0.025 m. 8. Air between transducer and skin would reflect almost all the ultrasound because their Z values are very different; gel has Z close to skin, so most is transmitted. 9. (0.37)² / (3.05)² = 0.015. 10. e^(−23 × 0.050) = 0.32. 11. A large fraction is reflected at the tissue-bone boundary and much of the rest is absorbed in bone, so little reaches the organ. 12. 2.5 cm (the same calculation as Question 7).