Cambridge A Level Physics (9702): Production and use of X-rays
A focused guide to 24.2 X-rays in Cambridge A Level Physics 9702: the X-ray tube, minimum wavelength, contrast, attenuation and how CT builds a 3D image.
- Subject
- Physics
- Level
- A LEVEL
- Topic
- Medical physics
- Author
- Marlbridge Academic Team
- Updated
- Reviewed by
- Hassan (what this means)
Aligned to Cambridge A Level Physics (9702), 2025-2027. Official specification .
Syllabus page (what it covers and how it is assessed): Cambridge A Level Physics.
Syllabus points this page covers
9702 (A Level)
- 24.2 Production and use of X-rays
Found an error? Report a correction.
Need help with this topic? Request a free trial class for A Level Physics (9702).
This guide covers subtopic 24.2 Production and use of X-rays of Cambridge International AS & A Level Physics 9702, 2025–2027 series (A Level). The broader Medical Physics study guide introduces all of Topic 24; this page goes further into how an X-ray tube works, why there is a minimum wavelength, how contrast arises and can be improved, how to compare beams that pass along different paths through the body, and the stages of a CT scan. For quick recall use the X-rays revision notes, and for exam-style questions the X-rays practice questions.
What 24.2 asks you to do
The syllabus lists four outcomes. Candidates should be able to:
- explain that X-rays are produced by electron bombardment of a metal target, and calculate the minimum wavelength of X-rays produced from the accelerating p.d.
- understand the use of X-rays in imaging internal body structures, including an understanding of the term contrast in X-ray imaging
- recall and use I = I₀e^(−μx) for the attenuation of X-rays in matter
- understand that computed tomography (CT) scanning produces a 3D image by combining X-ray images of one section taken from different angles into a 2D image of that section, then repeating along an axis and combining the 2D images of many sections
The X-ray tube
An X-ray tube is an evacuated glass or metal enclosure containing two electrodes.
- Cathode. A heated filament releases electrons by thermionic emission. The filament current controls how many electrons are released per second, and so the intensity of the X-ray beam.
- Accelerating p.d. A large p.d. V (tens of kilovolts up to more than 100 kV) between cathode and anode accelerates the electrons. Each electron gains kinetic energy eV.
- Target (anode). The electrons strike a metal target, often tungsten. As they decelerate rapidly in the target, some of their kinetic energy is emitted as X-ray photons.
- Vacuum. Prevents the electrons colliding with gas molecules on the way.
Only about 1 % of the electrons’ kinetic energy becomes X-rays; the rest becomes thermal energy in the target. That is why the target metal must have a high melting point and the anode is cooled, often by rotating it so the beam does not strike one spot continuously.
Tube current and power. With a tube current of 5.0 mA, the number of electrons striking the target each second is 5.0 × 10⁻³ / 1.60 × 10⁻¹⁹ = 3.1 × 10¹⁶. At 80 kV the power delivered to the target is VI = 80 000 × 5.0 × 10⁻³ = 400 W, nearly all of it as heat.
The minimum wavelength
An electron can lose its kinetic energy in one large step or several smaller ones, so the X-rays have a continuous range of photon energies. The largest photon energy occurs when an electron loses all its kinetic energy in a single interaction, producing one photon:
maximum photon energy = eV
hc / λ_min = eV so λ_min = hc / (eV)
Key consequences:
- λ_min depends only on the accelerating p.d., not on the target metal or the tube current.
- Doubling V halves λ_min.
- A higher p.d. produces more penetrating (harder) X-rays, with shorter wavelengths and higher photon energies.
Worked example 1: minimum wavelength
An X-ray tube operates at 100 kV. Calculate the maximum photon energy and the minimum wavelength.
E_max = eV = 1.60 × 10⁻¹⁹ × 100 × 10³ = 1.60 × 10⁻¹⁴ J (100 keV)
λ_min = hc / (eV) = (6.63 × 10⁻³⁴ × 3.00 × 10⁸) / 1.60 × 10⁻¹⁴
= 1.24 × 10⁻¹¹ m
Annotation. “100 kV” must become 100 × 10³ V. The answer is of order 10⁻¹¹ m, which is a useful check: diagnostic X-rays have wavelengths around 10⁻¹¹ m.
Attenuation in matter
As an X-ray beam passes through matter, photons are absorbed or scattered out of the beam, so its intensity falls exponentially:
I = I₀ e^(−μx)
μ is the linear attenuation coefficient (unit m⁻¹ when x is in metres). It depends on the material and on the photon energy. Bone has a much larger μ than soft tissue at diagnostic energies, mainly because it contains elements with higher proton numbers (such as calcium) and is denser.
Contrast
An X-ray image is a shadow picture: the detector (or film) records the intensity that passes through the body. Contrast is the difference in brightness (degree of blackening) between neighbouring areas of the image. Good contrast lets a doctor tell structures apart.
Contrast is good when adjacent structures have very different attenuation coefficients, such as bone next to soft tissue. It is poor between two soft tissues, whose μ values are similar.
Improving contrast
- Contrast media: a substance with a large attenuation coefficient is introduced into the body. A patient may swallow a barium compound to show the outline of the digestive tract, or have an iodine compound injected to show blood vessels.
- Choice of p.d.: lower-energy X-rays are attenuated more differently by different tissues, which can improve contrast between soft tissues; very high-energy X-rays pass through everything similarly. (The trade-off is that more low-energy radiation is absorbed by the patient.)
Worked example 2: comparing two paths
Illustrative values: μ(soft tissue) = 20 m⁻¹, μ(bone) = 60 m⁻¹. Beam A passes through 8.0 cm of soft tissue only. Beam B passes through 6.0 cm of soft tissue and 2.0 cm of bone. Compare the transmitted intensities.
A: I/I₀ = e^(−20 × 0.080) = e^(−1.60) = 0.20
B: I/I₀ = e^(−20 × 0.060) × e^(−60 × 0.020) = e^(−1.20 − 1.20) = e^(−2.40) = 0.091
ratio A : B = 0.20 / 0.091 = 2.2
Annotation. For layers of different materials, add the exponents (or multiply the factors). Twice as much intensity reaches the detector behind path A, so the bone appears clearly as a lighter shadow: good contrast. Two soft tissues of 5.0 cm with μ = 20 m⁻¹ and 22 m⁻¹ transmit 0.37 and 0.33: barely distinguishable.
Computed tomography (CT)
A single X-ray image flattens the whole body into one plane, so structures overlap. CT avoids this. Learn the stages in the syllabus order:
- An X-ray tube and a ring of detectors rotate around the patient, taking many X-ray images of the same thin section (slice) from different angles.
- A computer combines these images to produce a 2D image of that section.
- The patient is moved along an axis through the scanner and the process is repeated for the next section, and so on.
- The 2D images of multiple sections are combined to build a 3D image of the internal structure, which can be viewed from any angle or as any chosen slice.
As background beyond the syllabus outcome, a CT scan gives a much higher radiation dose than a single X-ray image, because the section is exposed from many directions.
Common mistakes
- Saying X-rays are produced when electrons “hit the target and bounce off”. They are produced as the electrons decelerate in the target.
- Using the wrong energy for λ_min: it is eV, the kinetic energy of one electron, not the power of the tube.
- Claiming λ_min depends on the target metal or the current.
- Defining contrast as “how sharp the image is”. Sharpness and contrast are different; contrast is the difference in blackening between regions.
- Multiplying μ values for two layers instead of adding μx terms.
- Describing CT as “a 3D X-ray” without the stages: many angles → 2D section → repeat along axis → 3D.
Exam technique for the command words
- Explain (how X-rays are produced): heated cathode, electrons accelerated by p.d., strike metal target, decelerate, kinetic energy emitted as X-ray photons; most becomes thermal energy.
- Explain (why there is a minimum wavelength): an electron transfers at most all its kinetic energy eV to one photon.
- Calculate λ_min: convert kV to V, then λ_min = hc/(eV).
- State (what is meant by contrast): the difference in degree of blackening between areas of the image.
- Suggest (how contrast may be improved): contrast medium with large μ, or suitable choice of X-ray energy.
- Describe (CT scanning): the four stages above, in order.
More resources on this topic
- Medical Physics study guide — the whole of Topic 24
- X-rays revision notes
- X-rays practice questions
- Production and use of ultrasound — subtopic 24.1
- PET scanning — subtopic 24.3
- Quantum Physics — photon energy E = hf = hc/λ
Written against Cambridge International AS & A Level Physics 9702, 2025–2027 series. Always check the current syllabus for your examination year.
Related resources
-
Practice questions
Cambridge A Level Physics (9702): Production and use of X-rays – Practice Questions
Original practice questions with mark-scheme answers on Cambridge A Level Physics 9702 24.2 X-rays: minimum wavelength, attenuation, contrast and CT scanning.
Physics · Cambridge · A Level
Try questions -
Revision notes
Cambridge A Level Physics (9702): Production and use of X-rays – Revision Notes
Revision notes for Cambridge A Level Physics 9702 24.2 X-rays: the X-ray tube, λ_min = hc/eV, contrast, attenuation and CT, with a self-test.
Physics · Cambridge · A Level